User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Kaye Instruments Digistrip 4S Recorder DATE: 28-Nov-95 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 10 NUMBER OF LINES: 173 CONFIGURATION: Fluke 5100B CONFIGURATION: HP 3458A STANDARD: Ectron 1120 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # Equipment #699-029 Equip. Type: KAYE Digistrip 4S Recorder # Notes:Due to needed accuracy, this equipment must be calibrated in the LAB. # D/693 Mark Williams # Review Approval:_____________________________ Release Date:____/___/____ # Revision Description. 1.001 ASK+ M X 1.002 ASK- R N P F W 1.003 STD Ectron 1120 1.004 DISP This procedure covers the Kaye Digistrip 4S 1.005 RSLT =Accuracy: 1 year specification 1.006 RSLT = 0.01% of reading + 1 LSD + 2æV 1.007 RSLT = 1.008 RSLT = 1.009 RSLT =Reference Documents: 1.010 RSLT = 1.011 HEAD Preliminary Setup 1.012 DISP The UUT and STANDARDS must have been on for 30-45 Min. 1.012 DISP prior to calibration to allow for stabilization. 1.013 HEAD mV RANGE FUNCTIONS 1.014 HEAD {Analog Board Calibration (mV INPUT)} 1.015 HEAD {6.5000V Range} 1.016 DISP Set UUT PROGRAM FUNCTION to 003 1.016 DISP and monitor channel 101 1.017 HEAD 6.0000V Reading 1.018 DISP Connect a set of copper leads from the 5100B 1.018 DISP to channel 101 of the UUT. 1.019 DISP Connect the 3458A DMM input to the 5100B output 1.020 IEEE [@3458]func,dcv,6,.00001 1.021 5100 6.00000V S 2W 1.022 3458 10.00000V N 2W 1.023 MEME 1.024 MEMI Enter the UUT indicated value. 1.025 MEME 1.026 TOL 0.01% 0.000102u 1.027 MEMC V TOL #! Test Tol 0.001102, Sys Tol 4.3e-005, TUR 25.628 (>= 4.00). 1.028 DISP Adjust P1 until UUT reads 6.0000 V 1.028 DISP with input of 6000mV from 5100 monitored by 3458A DMM. 2.001 5100 -6.00000V S 2W 2.002 3458 10.00000V N 2W 2.003 MEME 2.004 MEMI Enter the UUT indicated value. 2.005 MEME 2.006 TOL 0.01% 0.000102u 2.007 MEMC V TOL #! Test Tol 0.001102, Sys Tol 4.3e-005, TUR 25.628 (>= 4.00). 2.008 DISP Adjust P2 until UUT reads -6000.0 mV 2.008 DISP with input of -6000mV from 5100 monitored by 3458A DMM. 3.001 HEAD {650mV Range} 3.002 DISP Set UUT PROGRAM FUNCTION to 002 3.003 HEAD 600mV Reading 3.004 5100 600.000V S 2W 3.005 3458 1.00000V N 2W 3.006 MEME 3.007 MEMI Enter the UUT indicated mV value. 3.008 MATH mem=mem/1000 3.009 MEME 3.010 TOL 0.01% 0.000012u 3.011 MEMC mV TOL #! Test Tol 0.000112, Sys Tol 5.6e-006, TUR 20.000 (>= 4.00). 3.012 DISP Adjust P3 until UUT reads 600.0 mV 3.012 DISP with input of 600mV from 5100 monitored by 3458A DMM. 4.001 HEAD {65mV Range} 4.002 DISP Set UUT PROGRAM FUNCTION to 001 4.003 HEAD 60mV Reading 4.004 5100 0.06000V S 2W 4.005 3458 1.00000V N 2W 4.006 MEME 4.007 MEMI Enter the UUT indicated value. 4.008 MATH mem=mem/1000 4.009 MEME 4.010 TOL 0.01% 0.000003u 4.011 MEMC V TOL #! Test Tol 0.000103, Sys Tol 5.6e-006, TUR 18.393 (>= 4.00). 4.012 DISP Adjust P4 until UUT reads 60.0 mV 4.012 DISP with input of 60mV from 5100 monitored by 3458A DMM. 5.001 HEAD {30mV Range} 5.002 DISP Set UUT PROGRAM FUNCTION to 005 5.003 HEAD 6mV Reading 5.004 5100 0.03000V S 2W 5.005 3458 1.00000V N 2W 5.006 MEME >1 5.007 MEMI Enter the UUT indicated mV value... 5.008 MATH mem=mem/1000 5.009 MEME 5.010 MATH mem=m[1] 5.011 MATH mem=mem/5 5.012 TOL 0.01% 0.0000021u 5.013 MEMC mV TOL #! Test Tol 0.0001021, Sys Tol 5.6e-006, TUR 18.232 (>= 4.00). 5.014 DISP Adjust P5 until UUT reads 6.0 mV 5.014 DISP with input of 30mV from 5100 monitored by 3458A DMM. 6.001 HEAD {REFERENCE RTD Calibration; Type K} 6.002 DISP Set UUT PROGRAM FUNCTION to 027 6.003 DISP For accurate reference RTD calibration the UUT must 6.003 DISP have been ON for at least 2 hours. Ref pg 2-7 para 1 6.003 DISP of Digistrip III user's guide. 6.004 HEAD {Group 100 (32øF)} 6.005 DISP Connect a type K extension wire from the Ectron to 6.005 DISP UUT channel 101 6.006 DISP The UUT should already be monitoring channel 101. 6.007 DISP Slew the Ectron for a UUT reading of 6.007 DISP 32øF 6.008 ACC 32.000øF 0.018% 6.009 MEMI Enter the STANDARD indicated value. 6.010 MEMC øF 1U #! Test Tol 1, Sys Tol 0.00576, TUR 173.611 (>= 4.00). 6.011 DISP Adjust Group 100's reference junction pot P1 6.011 DISP until UUT reads 32øF 6.011 DISP with input of 32øF from Ectron. 7.001 HEAD {Group 200 (32øF)} 7.002 DISP Connect the type K extension wire to channel 201 7.003 DISP Set the UUT to monitor channel 201 7.004 DISP Slew the Ectron for a UUT reading of 7.004 DISP 32øF 7.005 ACC 32.000øF 0.018% 7.006 MEMI Enter the STANDARD indicated value. 7.007 MEMC øF 1U #! Test Tol 1, Sys Tol 0.00576, TUR 173.611 (>= 4.00). 7.008 DISP Adjust Group 200's reference junction pot P1 7.008 DISP until UUT reads 32øF 7.008 DISP with input of 32øF from Ectron. 8.001 HEAD {Group 300 (32øF)} 8.002 DISP Connect the type K extension wire to channel 301 8.003 DISP Set the UUT to monitor channel 301 8.004 DISP Slew the Ectron for a UUT reading of 8.004 DISP 32øF 8.005 ACC 32.000øF 0.018% 8.006 MEMI Enter the STANDARD indicated value. 8.007 MEMC øF 1U #! Test Tol 1, Sys Tol 0.00576, TUR 173.611 (>= 4.00). 8.008 DISP Adjust Group 300's reference junction pot P1 8.008 DISP until UUT reads 32øF 8.008 DISP with input of 32øF from Ectron. 9.001 HEAD {Group 400 (32øF)} 9.002 DISP Connect the type K extension wire to channel 401 9.003 DISP Set the UUT to monitor channel 401 9.004 DISP Slew the Ectron for a UUT reading of 9.004 DISP 32øF 9.005 ACC 32.000øF 0.018% 9.006 MEMI Enter the STANDARD indicated value. 9.007 MEMC øF 1U #! Test Tol 1, Sys Tol 0.00576, TUR 173.611 (>= 4.00). 9.008 DISP Adjust Group 400's reference junction pot P1 9.008 DISP until UUT reads 32øF 9.008 DISP with input of 32øF from Ectron. 10.001 DISP IF YOU HAVE MADE ANY ADJUSTMENTS, SAVE THESE RESULTS 10.001 DISP "AS FOUND". Cal due date is today 10.001 DISP and interval is BLANK. Rerun the procedure for your 10.001 DISP 2nd run and save results "AS LEFT". 10.002 HEAD 10.003 DISP ÿ P R O C E D U R E 10.003 DISP ÿ C O M P L E T E D ! #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.